Somatic Cell Count Tester Performance Verification
Compare with FOSS & Bently flow cytometry analyzer
Introduction
SCCT tester based on fluorescence microscopy technolgy
FOSS or Bently tester based on Flow CytoMetry technology
If SCCT tester’s result maximum close to FOSS or BENTLY analyzer’s result, that mean SCCT Tester has good performance and it’s reliable device for somatic cell count of raw milk.
Verification method
Mainly verify two factors: Correlation & Repeatability
International standard: ISO13366 (IDF148) Milk-Enumeration of somatic cells
Results verification: Compare with FOSS or Bently flo cytometry analyzer
Correlation verify
SCCT vs Bentley SCC analyzer

SCCT vs FOSS SCC analyzer

FOSS vs Bentley

Repeatability Verify
200,000cells/ml raw milk samples

500,000cells/ml raw milk samples

Conclusion
Correlation: > 0.98 (compare with FOSS and Bentley)
Repeatability:
200,000cells/ml <7%
500,000cells/ml <5%
SCCT Tester
1. Cost effective
2. Easy to use and maintainance
3. High accurate and reliable result
4. Low price per test
5. Portable and compact design
6. Worldwide distribution